by B. Zurro, A. Baciero, D. Jiménez-Rey, L. Rodríguez-Barquero and M. T. Crespo
Abstract:
We investigate the capability of a fast-ion luminescent probe to operate as a pulse height ion energy analyzer. An existing high sensitivity system has been reconfigured as a single channel ion detector with an amplifier to give a bandwidth comparable to the phosphor response time. A digital pulse processing method has been developed to determine pulse heights from the detector signal so as to obtain time-resolved information on the ion energy distribution of the plasma ions lost to the wall of the TJ-II stellarator. Finally, the potential of this approach for magnetic confined fusion plasmas is evaluated by studying representative TJ-II discharges.
Reference:
B. Zurro, A. Baciero, D. Jiménez-Rey, L. Rodríguez-Barquero and M. T. Crespo, “Investigating the performance of an ion luminescence probe as a multichannel fast-ion energy spectrometer using pulse height analysis”, Review of Scientific Instruments, vol. 83, no. 10, pp. 10D306.
Bibtex Entry:
@article{zurro_investigating_2012,
	title = {Investigating the performance of an ion luminescence probe as a multichannel fast-ion energy spectrometer using pulse height analysis},
	volume = {83},
	issn = {0034-6748},
	url = {http://aip.scitation.org/doi/10.1063/1.4733572},
	doi = {10.1063/1.4733572},
	abstract = {We investigate the capability of a fast-ion luminescent probe to operate as a pulse height ion energy analyzer. An existing high sensitivity system has been reconfigured as a single channel ion detector with an amplifier to give a bandwidth comparable to the phosphor response time. A digital pulse processing method has been developed to determine pulse heights from the detector signal so as to obtain time-resolved information on the ion energy distribution of the plasma ions lost to the wall of the TJ-II stellarator. Finally, the potential of this approach for magnetic confined fusion plasmas is evaluated by studying representative TJ-II discharges.},
	number = {10},
	urldate = {2017-10-23},
	journal = {Review of Scientific Instruments},
	author = {Zurro, B. and Baciero, A. and Jiménez-Rey, D. and Rodríguez-Barquero, L. and Crespo, M. T.},
	month = jul,
	year = {2012},
	pages = {10D306},
	file = {Full Text PDF:E:\cmam_papers\files\1237\Zurro et al. - 2012 - Investigating the performance of an ion luminescen.pdf:application/pdf;Full Text PDF:E:\cmam_papers\files\1252\Zurro et al. - 2012 - Investigating the performance of an ion luminescen.pdf:application/pdf;Full Text PDF:E:\Usuarios\Administrator\Zotero\storage\58XE4FNC\Zurro et al. - 2012 - Investigating the performance of an ion luminescen.pdf:application/pdf;Full Text PDF:E:\Usuarios\Administrator\Zotero\storage\S3DPFJ87\Zurro et al. - 2012 - Investigating the performance of an ion luminescen.pdf:application/pdf;Snapshot:E:\cmam_papers\files\1239\1.html:text/html;Snapshot:E:\cmam_papers\files\1254\1.html:text/html;Snapshot:E:\Usuarios\Administrator\Zotero\storage\6LSJBBNC\1.html:text/html;Snapshot:E:\Usuarios\Administrator\Zotero\storage\GVZLZ2G7\1.html:text/html},
}